Particulate contamination monitor for space launch applications
XCAM Scientific has spearheaded a technology development program funded by the European Space Agency (ESA) to create a unique piece of space technology equipment for measuring particulate fallout (PFO) inside a launch vehicle fairing during liftoff.
Concept of Operations
Central to the concept of operations, is the elimination of the need for tele-commanding between the launch vehicle (LV) and the instrument. When power is supplied to the unit, it automatically enters a pre-programmed power-up sequence before moving to a pre-flight data acquisition and transmission mode.
Using a combination of environmental sensors, the unit can autonomously detect when launch commences and move to the flight data acquisition and transmission mode, operating for a fixed duration between launch and fairing ejection. The unit then enters a post-flight transmission mode, transferring the remaining data to the launch vehicle for telemetry to the ground.
XCAM Scientific’s proprietary algorithms allow instant analysis of launch campaign data, detecting and classifying particles at any time.
Features & Benefits
- Ruggedised Design: ruggedised space technology equipment built to withstand severe vibration, shock, acoustic loads, rapid depressurisation, and extreme thermal environments experienced during liftoff.
- Standardised Mounting Points for Shock Absorbers: enables deployment across a wide range of launch vehicles.
- Surface Particle Monitoring from 5 μm: detecting and classifying particles and fibres every 10 s.
- Performance Verification: using a special chrome-on-glass test mask to confirm optimal operation.
- Space Qualified: the Space PFO is being considered for ESA’s Ariane-6 and Vega-C launch vehicles from 2027 launches onwards.
- Custom Development Available: to meet the demands of your mission
Why Monitor PFO?
Launch vehicles, and their enclosed satellite payloads, experience significant degrees of mechanical and acoustic shock and vibration during their journey into space.
Nothing is currently known about the sources and types of particulate contamination inside a rocket fairing during launch; however, it is certain that contamination can seriously affect the performance, accuracy and reliability of the multi-million dollar instruments which are routinely sent into space to study our planet and explore the universe.
Key Specifications
- Total Detection Area: 4 cm²
- Sample Rate: At least every 10 s
- Particle Readout: Size, shape, distribution; Particle/fibre classification
- Minimum Particle Size: 100% of 5 μm
- Operational Lifetime: Data acquisition: at least 254 s; Data transfer: at least 5 hours
- Operational Temperature: -20°C to +80°C
- Operational Humidity Range: 0% to 95%
- Instrument Volume / Mass: ~2.4 litres / 2.5 kg